Operator's Manual

Configuring the TM80/TM70 Maintenance Menu
TM80/TM70 Telemetry Monitor Operator’s Manual 13 - 3
3. Tap to exit the [General] menu.
Options Description Settings*
Display Auto Off Configures the time for display auto off. 1 min, 2 min, 5 min, 15 min 30
min, Off
Language Configures the system language. ENGLISH, FRENCH
Device Location Changes the hospital ,department name,
room number, and bed number.
Refer to "Configuring Device Location"
on page 13 - 4 for details.
N/A
Notch Filter Configures the ECG Notch filter. This
option is used to filter out AC line noise
from the ECG waveform.
50 Hz, 60 Hz, Off
The default is 50 Hz when the
monitor is not configured for the
US.
ECG Lead Labeling Changes the ECG lead labeling. AHA, IEC
Calibrate ECG Enables or disables the ECG verification. On, Off
Refer to "Verifying the ECG" on
page 17 - 5 for details.
SpO
2
Module Changes the SpO
2
module. Masimo SpO
2
, Nonin SpO
2
,
Nellcor SpO
2
SpO
2
Tone Configures the SpO
2
tone. Mode 1, Mode 2
Set Time/Date Set time and date for the TM80/TM70. Date &Time: 12Hr, 24 Hr
Date Format: DD/MM/YY, YY/MM/
DD, MM/DD/YY
Resp This option is available only when the
Resp functionality is supported.
When it is supported and [
Resp] is
enabled, Resp numeric and waveform are
displayed on the screen.
On, Off
* The factory default settings are in bold.
NOTE
Mindray recommends the same SpO
2
tone mode be used for the device
within a monitoring area.