Operator's Manual

Monitoring with the TM80/TM70 at the CMS SpO2 Monitoring
TM80/TM70 Telemetry Monitor Operator’s Manual 12 - 27
12.12.2 Changing SpO
2
Alarm Properties
To change ECG alarm properties, follow this procedure:
1. In the [SpO
2
] menu, select the [Alarm] tab and [PR Alarm] tab respectively.
2. Set the desired alarm properties.
For the default alarm properties, See."Factory Default Parameter Alarm Settings" on
page 12 - 3.
12.12.3 Changing SpO
2
Settings
1. In the [SpO
2
] menu, select the [SpO
2
Setup] tab.
2. Set the desired items.
12.12.4 SpO
2
Display
SpO
2
Digital Area
The SpO
2
digital area displays:
1. Parameter name
2. High SpO
2
alarm limit
3. PR label
4. PR value
5. High PR alarm limit
6. SpO
2
unit of measure
7. Low PR alarm limit
8. Masimo perfusion index value
9. PI label
Options Description Settings*
Sensitivity
(Masimo only)
The option is not configurable. The option
setting is synchronous with the setting
at the TM80/TM70. See
"Configuring
the SpO2 Setup" on page 8 - 6
".
None
Display PI
Speed Configures the Pleth wave speed. 25 mm/s
Averaging
(Masimo only)
The option is not configurable. The option setting is synchronous with the set-
ting at the TM80/TM70. See "Configuring the SpO2 Setup" on page 8 - 6 for
details.