Monitoring with the TM80/TM70 at the CMS Arrhythmia Monitoring
TM80/TM70 Telemetry Monitor Operator’s Manual 12 - 21
4. Baseline ST value
5. ST unit
5-lead ST segments
12.10 Arrhythmia Monitoring
12.10.1 Entering the Arrhythmia Tab
To enter the [Arrhythmia] tab, follow this procedure:
1. Select the ECG digital area or waveform area to enter the [ECG] menu on the
ViewBed screen of the CMS.
2. Select the [Arrhythmia] tab.
12.10.2 Measurement Limitations
■ Since the arrhythmia detection algorithm sensitivity and specificity is less than
100%, sometimes there may be some false arrhythmias detected and also some
true arrhythmia events may not be detected. This is especially true when the
signal is noisy.
■ The ECG size and QRS width settings affect arrhythmia detection and heart rate
calculation sensitivity.
■ If QRS amplitude is low, the monitor might not be able to calculate HR and false
asystole may occur.
■ At the algorithm learning phase, arrhythmia detection may not be available.
Therefore, closely monitor the patient’s condition during the learning phase after
the learning phase to allow the algorithm to reach optimal detection
performance.