Operator's Manual
ST Monitoring Monitoring with the TM80/TM70 at the CMS
12 - 20 TM80/TM70 Telemetry Monitor Operator’s Manual
4. Set the desired items.
12.9.8 ST Display
When [ST Analysis] is enabled, the ST digital area displays:
1. Parameter label
2. ST unit
3. ST alarm off symbol
4. Lead labels
5. ST numerics: a positive value indicates ST segment elevation, and a negative value
indicates ST segment depression.
Example 5-lead ST digital area
12.9.9 ST Segment Display
The ST segments display in the waveform area:
1. Lead labels
2. ST markers (ISO, J/ST)
3. Current ST value
Buttons Description Settings*
Set Baseline Set an ST baseline when ST values become
stable.
None
The QT template updated time displays at the bottom of the screen.
Display/Hide Baseline Displays or hides the reference baseline. None
Display/Hide Marker Displays or hides the position of ISO point,
J point and ST point.
1
3
4
5
2