Operator's Manual
Monitoring with the TM80/TM70 at the CMS QT Monitoring
TM80/TM70 Telemetry Monitor Operator’s Manual 12 - 13
12.8.2 Entering the QT Tab
To enter the [QT] tab, follow this procedure:
1. Select the ECG digital area or waveform area to enter the [ECG] menu on the
ViewBed screen of the CMS.
2. Select the [QT] tab.
12.8.3 Changing QT Alarm Properties
To set QT alarm properties, follow this procedure:
1. From the [QT] tab, select the [Alarm] section.
2. Set the desired alarm properties.
For the default alarm properties, See "Factory Default Parameter Alarm Settings" on
page 12 - 3.
12.8.4 Changing QT Settings
To change QT related settings, follow this procedure:
1. From the [QT] tab, select the [Setup] section.
2. Set the desired items.
12.8.5 QT View Menu
Select [QT View] at the bottom of the [QT] menu to enter the [QT View] menu.
In the [QT View] menu, you can view a snapshot of the real-time wave and to verify that
the QT algorithm detects correct Q and T points.
The [QT View] menu displays, as shown in the following figure:
■ The current waveform and parameter values display in green.
Options Description Default Setting
QT Analysis Enables or disables QT analysis. Off
QT View Select to display the QT View
menu.
None
This button is only available for the advanced ECG algorithm. See "QT View Menu" on
page 12 - 13 for details.
Set Baseline Configures the baseline to quan-
tify changes in the QTc value.
None