Operator's Manual
Arrhythmia Monitoring Monitoring ECG
7 - 24 TM80/TM70 Telemetry Monitor Operator’s Manual
4. Tap to return to the previous menu.
7.9 Arrhythmia Monitoring
Arrhythmia monitoring is intended for adult and pediatric patients.
7.9.1 Arrhythmia Safety Information
Delta QTc Configures whether to trigger
the Delta QTc alarm.
On, Off
Upper alarm limit range: 30ms to 200 ms; the
step is 1ms
The default is 60 ms.
Alarm priority: Low, Med, High
Options Description Settings*
* The factory default settings are in bold.
WARNING
• Heart rate reading may be affected by cardiac arrhythmias. Do not rely
entirely on heart rate alarms when monitoring patients with arrhythmia.
Always keep these patients under close surveillance.
• The arrhythmia analysis program is intended to detect ventricular
arrhythmias and atrial fibrillation. It is not designed to detect atrial or
supraventricular arrhythmias. It may incorrectly identify the presence or
absence of an arrhythmia. Therefore, a physician must analyze the
arrhythmia information with other clinical findings.