Operator's Manual
Monitoring with the TM80/TM70 at the CMS ECG Monitoring
TM80/TM70 Telemetry Monitor Operator’s Manual 12 - 9
12.7.1 Changing ECG Alarm Properties
1. From the [ECG] tab, select the [Alarm] section.
2. Set the desired alarm properties.
For the default alarm properties, See "Factory Default Parameter Alarm Settings" on
page 12 - 3.
12.7.2 Changing ECG Wave Settings
1. From the [ECG] tab, select the [Setup] section.
2. Set the desired items.
Options or Buttons Description Default Setting
ECG 1 Select the desired ECG lead and set the
corresponding gain for ECG 1.
II
ECG 2 (5-lead, 6-lead) I
Va (For 6-lead only) Va
Vb (For 6-lead only) Vb
ECG Gain Configures the size of each ECG waveform. x1
Speed Configures the ECG wave speed. 25 mm/s
Filter Configures the ECG filter in all operating
modes.
■ [Monitor]: use under normal
measurement conditions.
■ [ST]: use when the ST monitoring
is applied.
Monitor
Notch Filter Configures whether or not to switch on
the notch filter.
On
Lead Set 3-Lead, 5-Lead, 6-Lead
* The factory default settings are in bold.