Operator's Manual

Arrhythmia Threshold Settings Monitoring with the TM80/TM70 at the CMS
12 - 8 TM80/TM70 Telemetry Monitor Operator’s Manual
12.6 Arrhythmia Threshold Settings
When an arrhythmia violates its threshold, an alarm is triggered. When HR is less than 30
bpm, it is recommended to set asystole delay to 10 s.
12.7 ECG Monitoring
At the CMS, you can view and change the ECG settings in the [ECG] tab.
To enter the [ECG] tab, follow this procedure:
1. Select the ECG digital area or waveform area to enter the [ECG] menu on the
ViewBed screen of the CMS.
2. Select the [ECG] tab.
Arrh. event Range or Option Default Step
Unit of
measure
PVCs /min 1 to 100 10 1 minute
Pauses/min 1 to 15 8 1 None
Asystole Delay 3 to 10 4 1 second
Tachy(HRHigh) 60 to 300 Adult: 120
Pediatric: 160
5bpm
Brady(HR Low) Adult: 15 to 115
Pediatric: 15 to 120
Adult: 50
Pediatric: 75
5bpm
Extreme Tachy 60 to 300 Adult: 160
Pediatric: 180
5bpm
Extreme Brady Adult: 15 to 115
Pediatric: 15 to 120
Adult: 35
Pediatric: 50
5bpm
Multif PVC’ Window 3 to 31 15 1 beat
V-Tac Rate 100 to 200 130 5 bpm
V-Tac PVC 3 to 99 6 1 beat
Pause Threshold 1.5, 2, and 2.5 2 0.5 second
V Brady Rate 15 to 60 40 5 bpm
V Brady PVCs 3 to 99 5 1 beat