Monitoring with the TM80/TM70 at the CMS Factory Default Arrhythmia Alarm Settings
TM80/TM70 Telemetry Monitor Operator’s Manual 12 - 7
In addition, the activation state of some arrhythmias can be set as a whole with the fol-
lowing buttons that are at the bottom of the [Arrhythmia Alarms] tab.
Missed Beats On, Off High, Med, Low, Prompt On, Off
Multiform PVC High, Med, Low, Prompt
Nonsus V-Tac
Pause High, Med, Low, Prompt
A-Fib High, Med, Low, Prompt
Irr Rhythm Low, Prompt
Pauses/min High, Med, Low, Message
Name Activation State Alarm Priority
Alarm
Outputs
Notes
The factory default settings are in bold.
NOTE
• When Paced is set to Yes, the Missed Beat (MIS) alarm is reported as the
Pacer Not Capture (PNC) or Pacer Not Pacing (PNP) alarm.
• The priority of lethal arrhythmia alarms is always high. It is unchange-
able.
Button Description
Lethals Only Sets the lethal arrhythmia alarms to on and all non-lethal arrhythmia alarms to off.
All On Sets all arrhythmia alarms to on.
All Off Sets all arrhythmia alarms to off.