Operator's Manual

Monitoring with the TM80/TM70 at the CMS Factory Default Arrhythmia Alarm Settings
TM80/TM70 Telemetry Monitor Operator’s Manual 12 - 7
In addition, the activation state of some arrhythmias can be set as a whole with the fol-
lowing buttons that are at the bottom of the [Arrhythmia Alarms] tab.
Missed Beats On, Off High, Med, Low, Prompt On, Off
Multiform PVC High, Med, Low, Prompt
Nonsus V-Tac
Pause High, Med, Low, Prompt
A-Fib High, Med, Low, Prompt
Irr Rhythm Low, Prompt
Pauses/min High, Med, Low, Message
Name Activation State Alarm Priority
Alarm
Outputs
Notes
The factory default settings are in bold.
NOTE
When Paced is set to Yes, the Missed Beat (MIS) alarm is reported as the
Pacer Not Capture (PNC) or Pacer Not Pacing (PNP) alarm.
The priority of lethal arrhythmia alarms is always high. It is unchange-
able.
Button Description
Lethals Only Sets the lethal arrhythmia alarms to on and all non-lethal arrhythmia alarms to off.
All On Sets all arrhythmia alarms to on.
All Off Sets all arrhythmia alarms to off.