Operator's Manual

Monitoring with the TM80/TM70 at the CMS Factory Default Parameter Alarm Settings
TM80/TM70 Telemetry Monitor Operator’s Manual 12 - 5
SpO
2
Unit of mea-
sure: %
Invalid data:
“--”
On, Off For adult and
pediatric:
(Low limit + 2)
to 100
Default: 100
For adult
and pediat-
ric:
0 to (high
limit - 2)
Default: 90
High,
Med
On, Off
SpO
2
Desat
Unit of mea-
sure: %
Invalid data:
“--”
For adult and
pediatric:
None
For adult
and pediat-
ric:
0 to 100
Default: 80
High
NIBP-S
Unit of mea-
sure: mmHg
Invalid data:
“---”
For adult:
(Low limit + 5)
to 270
Default: 160
For adult:
40 to (high
limit - 5)
Default: 90
High,
Med
NIBP-D
Unit of mea-
sure: mmHg
Invalid data:
“---”
For adult:
(Low limit + 5)
to 210
Default: 90
For adult:
10 to (high
limit - 5)
Default: 50
NIBP-M
Unit of mea-
sure: mmHg
Invalid data:
“---”
For adult:
(Low limit + 5)
to 230
Default: 110
For adult:
20 to (high
limit - 5)
Default: 60
Parameters
Activati
on
State
Alarm High
Limit
Alarm Low
Limit
Alarm
Priority
Alarm Outputs
Notes:
ΔQTc is only available for Advanced ECG Algorithm.
The factory default settings are in bold.
NOTE
The SpO
2
Desat alarm limit is restricted such that the alarm limit can not
be higher than the SpO
2
low limit.