Operator's Manual

Factory Default Parameter Alarm Settings Monitoring with the TM80/TM70 at the CMS
12 - 4 TM80/TM70 Telemetry Monitor Operator’s Manual
Parameters
Activati
on
State
Alarm High
Limit
Alarm Low
Limit
Alarm
Priority
Alarm Outputs
HR/PR
Unit of mea-
sure: bpm
Invalid data:
“--”
On, Off For adult:
(Low limit + 2)
to 300
Default: 120
For pediatric:
(Low limit + 2)
to 350)
Default: 160
For adult:
15 to (high
limit - 2)
Default: 50
For pediat-
ric:
15 to (high
limit - 2)
Default: 75
High,
Med
On, Off
ST Single
Unit of mea-
sure: mV
Invalid data:
“--”
On, Off For adult and
pediatric:
(Low limit +
0.2) to 2.00
Default: 0.1
For adult
and pediat-
ric:
-2.00 to
(high limit –
0.2)
Default: -0.1
High,
Med,
Low
ST Dual
Unit of mea-
sure: mV
Invalid data:
“--”
QTc
Unit of mea-
sure: ms
Invalid data:
“--”
For adult:
200 to 800
Default: 500
For pediatric:
200 to 800
Default: 480
For adult
and pediat-
ric: None
ΔQTc*
Unit of mea-
sure: ms
Invalid data:
“--”
For adult and
pediatric:
30 to 200
Default: 60
For adult
and pediat-
ric: None
RR
Unit of mea-
sure: rpm
Invalid data:
“--”
On, Off
For adult and
pediatric:
10 to 100
Default: 30
For adult
and pediat-
ric: 0 to 55
Default: 8
Notes:
ΔQTc is only available for Advanced ECG Algorithm.
The factory default settings are in bold.