Operator's Manual
Introduction Monitoring with the TM80/TM70 at the CMS
12 - 2 TM80/TM70 Telemetry Monitor Operator’s Manual
12.1 Introduction
The chapter describes detailed configurations and data displayed on the BeneVision
CMS once the TM80/TM70 are connected to the CMS.
When the TM80/TM70 is successfully connected to the CMS, the TM80/TM70 provides
the following functions.
■ The TM80/TM70 can transmits parameter valu
es, waveforms, alarm settings, and
events to the CMS. From the CMS, you can check the patient’s monitoring data
and alarms.
■ Patient information, alarm settings, and alarm status can be synchronized
between the TM80/TM70 and the CMS.
■ If the TM80 is temporarily disconnected from the CMS, it can resend up data of the
last two hours once it is reconnected to the CMS while it can only resend data of
the last 10 second for TM70 when it’s disconnected and then reconnected to the
CMS.
■ The TM80/TM70’s screen is off.
12.2 Safety
CAUTION
• Wireless network designing, deploying, debugging, and maintenance
should be executed by Mindray service personnel or authorized techni-
cians.
• Always set the wireless network according to local wireless regulations.
• Keep network authentication information, for example password, safe,
protecting the network from being accessed by unauthorized users.
• Do not connect non-medical devices to the monitor network.
• If wireless network signal is poor, there may be a risk of CMS data loss.
• RF interference may result in wireless network disconnection.
• Disconnecting from the network may result in CMS data loss and func-
tion failure. Check the patient in case of network disconnection and
solve the network problem as soon as possible.
• Ensure that the monitor IP address setting is correct. Changing the net-
work settings may result in network disconnection. Contact your service
personnel if you have any problems on setting the IP address.