Interactions after Pairing a TM80/TM70 with a BP10 Monitoring NIBP (Optional)
10 - 10 TM80/TM70 Telemetry Monitor Operator’s Manual
DIA
Extreme
Configures whether to trigger
the extreme NIBP diastolic
pressure alarm.
On, Off
Alarm limit range:
For adult:
10 mmHg to 210 mmHg;
the default alarm upper limit is 105 mmHg
and the default alarm low limit is 35 mmHg
For pediatric:
10 mmHg to 150 mmHg; the default alarm
upper limit is 80 mmHg and the default alarm
low limit is 30mmHg
Alarm priority: High
MAP
Extreme
Configures whether to trigger
the extreme NIBP mean
pressure alarm.
On, Off
Alarm limit range:
For adult:
20 mmHg to 230 mmHg;
the default alarm upper limit is 125 mmHg
and the default alarm low limit is 45 mmHg
For pediatric:
20 mmHg to 165 mmHg; the default alarm
upper limit is 100 mmHg and the default
alarm low limit is 40mmHg
Alarm priority: High
Options Description Settings*
* The factory default settings are in bold.
NOTE
The upper alarm limit of systolic pressure, diastolic pressure, and mean
pressure cannot be greater than that of extreme systolic pressure,
extreme diastolic pressure, and extreme mean pressure.
The lower alarm limit of systolic pressure, diastolic pressure, and mean
pressure cannot be less than that of extreme systolic pressure, extreme
diastolic pressure, and extreme mean pressure.