Operator's Manual

Monitoring NIBP (Optional) Interactions after Pairing a TM80/TM70 with a BP10
TM80/TM70 Telemetry Monitor Operator’s Manual 10 - 9
DIA Configures whether to trigger
the NIBP diastolic pressure
alarm.
On, Off
Alarm limit range:
For adult:
11 mmHg to 209 mmHg; the default alarm
upper limit is 90 mmHg and the default alarm
low limit is 50mmHg
For pediatric:
11 mmHg to 149 mmHg; the default alarm
upper limit is 70 mmHg and the default alarm
low limit is 40mmHg
Alarm priority: Med, High
MAP Configures whether to trigger
the NIBP mean pressure alarm.
On, Off
Alarm limit range:
For adult:
21 mmHg to 229mmHg; the default alarm
upper limit is 110 mmHg and the default
alarm low limit is 60mmHg
For pediatric:
Alarm limit range: 21 mmHg to 164 mmHg;
the default alarm upper limit is 90 mmHg and
the default alarm low limit is 50mmHg
Alarm priority: Med, High
SYS
Extreme
Configures whether to trigger
the extreme NIBP systolic
pressure alarm.
On, Off
Alarm limit range:
For adult:
40 mmHg to 270 mmHg;
the default alarm upper limit is 175 mmHg
and the default alarm low limit is 75 mmHg
For pediatric:
40 mmHg to 200 mmHg; the default alarm
upper limit is 130 mmHg and the default
alarm low limit is 60mmHg
Alarm priority: High
Options Description Settings*
* The factory default settings are in bold.