Operator's Manual
Interactions after Pairing a TM80/TM70 with a BP10 Monitoring NIBP (Optional)
10 - 8 TM80/TM70 Telemetry Monitor Operator’s Manual
10.5.3 NIBP Operations at the TM80/TM70
In the [Setup] section of the [NIBP] menu at the TM80/TM70, you can:
■ Tap [Mode] to set NIBP measurement mode to one of the following: Manual, Auto,
Seq. ABPM.
■ Tap [ABPM] to configure ABPM-related settings: Day, Night, Start, Interval.
■ Tap[ Auto] to set NIBP measurement interval to one of the following: 1 min, 2 min,
2.5 min, 3 min, 5 min, 10 min, 15 min,20 min, 30 min, 1h, 1.5h, 2h, 3h, 4h, 8h.
■ Tap [Seq.] to configure sequence-related settings: P1, P2, P3, P4, P5
■ Tap [Initial Pressure] to set the initial cuff pressure.
■ Tap [Color] to select your desired color in which NIBP measurement results will be
displayed on the main screen of the TM80/TM70. 16 color options are available.
Tap your desired color and this color will be displayed on the right of [Color].
For detailed description of NIBP-related operations and settings, refer to BP10 NIBP
Module Operator’s Manual (P/N 046-008269-00).
10.5.4 Adjusting NIBP Limits
1. In the [Alarms] section of the [NIBP] menu, tap [NIBP Alarm Setup] .
2. Configure the options described in the following table.
Options Description Settings*
SYS Configures whether to trigger
the NIBP systolic pressure alarm.
On, Off
Alarm limit range:
For adult:
41 mmHg to 269 mmHg; the default alarm
upper limit is 160 mmHg and the default
alarm low limit is 90mmHg
For pediatric:
41 mmHg to 199 mmHg; the default alarm
upper limit is 120 mmHg and the default
alarm low limit is 70mmHg
Alarm priority: Med, High
* The factory default settings are in bold.