Operator's Manual

Unpairing the TM80/TM70 with the BP10 Monitoring NIBP (Optional)
10 - 4 TM80/TM70 Telemetry Monitor Operator’s Manual
case, follow steps in Pairing the TM80/TM70 with the BP10 on Page 10-2 to
pair the TM80/TM70 with the BP10 again.
When NIBP measurement mode is set to [Manual], [Auto], or [Sequence] and the
BP10 is paired successfully with the TM80/TM70, the wireless setup confirmation
message “Select Yes to continue Patient in BP10. Select No to continue Patient in
Telemetry.” will not be displayed. Patient-related settings from the TM80/TM70 will
be transferred to the BP10. But NIBP-related settings displayed on the main screen
of the TM80/TM70 will come from the BP10. The patient monitored by BP10 will be
discharged automatically and the data history related to this patient will be
cleared.
Patient Category can be changed at the TM80/TM70 only.
A patient cannot be discharged from the BP10 and NIBP-related settings are
grayed out at BP10. To discharge a patient from BP10, you need to unpair the BP10
form the TM80/TM70, refer to "Unpairing the TM80/TM70 with the BP10" on
page 10 - 4.
10.3 Unpairing the TM80/TM70 with the BP10
You can unpair the TM80/TM70 with the BP10 either via the TM80/TM70 or via the BP10.
10.3.1 Unpairing via the TM80/TM70
Follow this procedure to unpair the TM80/TM70 with the BP10 via the TM80/TM70.
1. Press to enter the main menu of the TM80/TM70.
2. Tap [Wireless Modules].
3. Select the device you wish to disconnect from the TM80/TM70 from the devices
list under [Devices] and tap [Disconnect]. The connected device is disconnected
from the TM80/TM70.
You can also follow this procedure to unpair it with the BP10.
1. Press the key to enter the main menu of the TM80/TM70.
2. Tap [Wireless Modules].
3. Tap [MPAN].
4. Select [Off] or [Sensor].