Operator's Manual
Unpairing the TM80/TM70 with the BP10 Monitoring NIBP (Optional)
10 - 4 TM80/TM70 Telemetry Monitor Operator’s Manual
case, follow steps in Pairing the TM80/TM70 with the BP10 on Page 10-2 to
pair the TM80/TM70 with the BP10 again.
■ When NIBP measurement mode is set to [Manual], [Auto], or [Sequence] and the
BP10 is paired successfully with the TM80/TM70, the wireless setup confirmation
message “Select Yes to continue Patient in BP10. Select No to continue Patient in
Telemetry.” will not be displayed. Patient-related settings from the TM80/TM70 will
be transferred to the BP10. But NIBP-related settings displayed on the main screen
of the TM80/TM70 will come from the BP10. The patient monitored by BP10 will be
discharged automatically and the data history related to this patient will be
cleared.
■ Patient Category can be changed at the TM80/TM70 only.
■ A patient cannot be discharged from the BP10 and NIBP-related settings are
grayed out at BP10. To discharge a patient from BP10, you need to unpair the BP10
form the TM80/TM70, refer to "Unpairing the TM80/TM70 with the BP10" on
page 10 - 4.
10.3 Unpairing the TM80/TM70 with the BP10
You can unpair the TM80/TM70 with the BP10 either via the TM80/TM70 or via the BP10.
10.3.1 Unpairing via the TM80/TM70
Follow this procedure to unpair the TM80/TM70 with the BP10 via the TM80/TM70.
1. Press to enter the main menu of the TM80/TM70.
2. Tap [Wireless Modules].
3. Select the device you wish to disconnect from the TM80/TM70 from the devices
list under [Devices] and tap [Disconnect]. The connected device is disconnected
from the TM80/TM70.
You can also follow this procedure to unpair it with the BP10.
1. Press the key to enter the main menu of the TM80/TM70.
2. Tap [Wireless Modules].
3. Tap [MPAN].
4. Select [Off] or [Sensor].