Operator's Manual

Monitoring NIBP (Optional) Pairing the TM80/TM70 with the BP10
TM80/TM70 Telemetry Monitor Operator’s Manual 10 - 3
T.
10.2.2 System Responses after Successful Pairing
Once a TM80/TM70 is paired successfully with a BP10, the following system responses
occur:
The icon will be changed to in the top-right corner of the TM80/TM70’s
screen. Besides, the appears in the top-right corner of the BP10.
The status of the BP10 which is paired successfully with the TM80/TM70 is
displayed as [Connected]. Besides, the “Pairing Successful” displays in the BP10’s
message area for about 10 seconds.
When NIBP measurement mode is set to [ABPM] at BP10 and the BP10 is paired
successfully with the TM80/TM70, the wireless seup confirmation message “Select
Yes to continue Patient in BP10. Select No to continue Patient in Telemetry.” will
display on the main screen of the TM80/TM70.
[Ye s ]: Patient-related settings comes from BP10 and ongoing NIBP
measurement for current patient will not stop.
[No]: Patient-related settings at BP10 comes from the TM80/TM70. Besides,
NIBP measurement for the patient monitored by BP10 will stop and the data
history related to this patient will be cleared at BP10.
If neither [Ye s ] nor [No] is selected, the wireless setup confirmation message
will disappear in three minutes. Besides, the “Pairing failure” message will be
displayed in the prompt message area of the TM80/TM70 . The “MPAN
Disconnected” message is displayed in the message area of the BP10. In this
WARNING
Do not pair a TM80/TM70 with an undesired BP10. If a TM80/TM70 is
paired with an undesired BP10, the patient category may be not applica-
ble for the patient and the measured NIBP data may be incorrect.
NOTE
Make sure that you have selected the correct BP10 to pair it with the
TM80/TM70.
If you pair a TM80/TM70 with a BP10 when NIBP measurement mode is
set to [ABPM] at BP10, the system responses after successful pairing are
slightly different from that when the NIBP measurement mode is set to
[Manual], [Auto], or [Sequence]. For the differences, see "System