Operator's Manual
Introduction Monitoring NIBP (Optional)
10 - 2 TM80/TM70 Telemetry Monitor Operator’s Manual
10.1 Introduction
NIBP (non-invasive blood pressure) data is measured by BP10 using the oscillometric
method. This measurement can be used for adult and pediatric patients. NIBP data is
transferred from the BP10 to the TM80/TM70.
You can pair and unpair a TM80/TM70 with a BP10 via the Mindray Patient Area Network
(MPAN) functionality.
Once a TM80/TM70 is paired successfully with a BP10, measured NIBP data and NIBP-
related alarms will be displayed on the TM80/TM70 main screen. Besides, the TM80/
TM70 allows you to perform some operations on the BP10.
For the functionalities and operations of the BP10, refer to BP10 NIBP Module Opera-
tor’s Manual (P/N 046-008269-00).
For the specifications of MPAN, refer to "Bluetooth Specification" on page A - 12.
10.2 Pairing the TM80/TM70 with the BP10
10.2.1 Pairing Procedure
Before pairing the TM80/TM70 with a new BP10, first unpair the already connected BP10
by following steps in "Unpairing the TM80/TM70 with the BP10" on page 10 - 4.
To pair the TM80/TM70 with the BP10, follow this procedure:
1. Press the MPAN key on the right panel of the BP10. The “Pairing...” message will dis-
play in the message area of the BP10.
2. Press to enter the main menu of the TM80/TM70.
3. Tap [Wireless Modules].
4. Tap [MPAN], and then select [Monitor/Sensor].
The TM80/TM70 starts searching the BP10 devices in the vicinity and displays the
devices that can be connected under [Devices]. Make sure that the device you
wish to connect appears in the list of devices. If not, repeat Steps 1 to 4.
5. In the list of devices, select your desired device and tap [Connect]. Once the
device is connected successfully, the status will be changed from [Disconnect] to
[Connected].
6. Start an NIBP measurement from TM80/TM70 to verify that desired BP10 module is
paired, and the NIBP measurement displayed on the BP10 are consistent with that
displayed on the TM80/TM70.