Operator's Manual
Monitoring Resp (Optional) Changing Resp Settings
TM80/TM70 Telemetry Monitor Operator’s Manual 9 - 5
9.4.2 Configuring the Resp Waveform
1. In the [Waveform] section of the [Resp] menu, select the options described in the
following table.
2. Press to return to the main screen.
9.4.3 Configuring the Resp Alarm Settings
1. In the [Alarms] section of the [Resp] menu, tap [Resp Alarm Setup] .
2. Select the options described in the following table.
Options Description Settings*
Gain Selects the Resp waveform size. ×0.25, ×0.5, ×1, ×2, ×3, ×4,
×5
Speed Selects the Resp waveform speed. 3 mm/s, 6.25 mm/s, 12.5
mm/s, 25 mm/s
* The factory default settings are in bold.
Options Description Settings*
RR Configures whether to switch on the Respiration
alarm. RR defaults to be switched on.
After tapping RR, you can configure alarm limits
and alarm priority.
On, Off
Alarm limit range:
0 RPM to 100 RPM
The default alarm upper
limit is 30 RPM. The default
alarm lower limit is 8 RPM.
Alarm Priority: Low, Med,
High
No Chest Movement The TM80/TM70 will trigger the “No Chest Move-
ment” alarm if the patient has stopped breath-
ing for longer than the set apnea time. This
option defaults to be switched on.
When the TM80/TM70 is connected to the
Central Monitoring System, switching on or off
this option at the Central Monitoring System
also switches it on or off at the TM80/TM70, and
vice versa.
After tapping this option, you can configure the
alarm priority.
Alarm Priority: High
* The factory default settings are in bold.