Operator's Manual

Changing Resp Settings Monitoring Resp (Optional)
9 - 4 TM80/TM70 Telemetry Monitor Operator’s Manual
9.4 Changing Resp Settings
You can change the Resp settings from the [Resp] menu.
9.4.1 Configuring the Resp Setup
Enter the [Resp] menu in either of the following ways:
On the main screen, tap the Resp digital area or Resp waveform area to enter the
[Resp] menu.
In the main menu, tap [Parameter Setup] [Resp] to enter the [Resp] menu.
1. In the [Setup] section of the [Resp] menu, select the options described in the fol-
lowing table.
2. Press to return to the main screen.
NOTE
Store the electrodes at room temperature. Open the electrode package
immediately prior to use.
Check that the electrode packages are intact and not expired. Make sure
the electrode gel is moist.
Options Description Settings*
NCM Alarm Delay
Time
Select the delay for no chest movement alarm. 10 s, 15 s, 20 s, 25 s, 30 s, 35
s, 40 s
Color Select the color of RR value and Resp waveform.S 16 colors
The default color is yellow.
* The factory default settings are in bold.