ST Monitoring Monitoring ECG
7 - 20 TM80/TM70 Telemetry Monitor Operator’s Manual
7.7.3 Displaying ST Numerics
When [ST Analysis] is switched on, ST digital area is displayed on the screen. The loca-
tion of the ST digital area is dependent on the display configurations. For how to set the
location of the ST digital area, refer to "Configuring the Display" on page 4 - 2.
Numerics displayed in the ST digital area are different according to the lead type:
■ When the 3-lead ECG leadwires are used, the ST digital area is not displayed. A ST
value is displayed in the HR digital area.
■ When the 5-lead ECG leadwires are used, seven ST values (ST-I, ST-II, ST-III, ST-aVR,
ST-aVL, ST-aVF, ST-V) are displayed in the ST digital area.
■ When the 6-lead ECG leadwires are used, eight ST values (ST-I, ST-II, ST-III, ST-aVR,
ST-aVL, ST-aVF, ST-V
a, ST-Vb) are displayed in the ST digital area.
This example shows the ST digital area when the 6-lead ECG leadwires are used. Your
screen may look slightly different:
7.7.4 Setting ST Alarms
1. Tap the ST digital area to enter the [ST] menu.
2. Tap [ST Alarm Setup].
3. In the [Setup] section, tap [ST Alarm Mode] and then select the desired mode:
◆ [
Absolute]: you can separately set the alarm properties for each ST
alarm. After selecting this option, the [
Auto Limits] option is displayed
where you can initiate auto adjustment of alarm limits.
◆ [Relative]: it is the default option. You can set the alarm properties for [ST
Single] and [ST Dual] alarms.
1. Parameter label 2. ST alarm activation state off icon
3. Lead label 4. ST unit
5. ST numerics: a positive value indicates ST segment elevation, and a negative value indi-
cates ST segment depression.
1
3
4
)
2
5