Operator's Manual
Monitoring SpO2 (Optional) Changing the SpO2 Settings
TM80/TM70 Telemetry Monitor Operator’s Manual 8 - 9
8.5.2 Configuring the SpO
2
Waveform
1. In the [Waveform] section of the [SpO
2
] menu, select the options described in the
following table.
2. Tap to exit the [SpO2] menu.
8.5.3 Configuring SpO
2
Alarm Settings
You can change the SpO
2
and SpO
2
Desat alarm properties. You can also set whether to
measure SpO
2
and NIBP simultaneously.
1. In the [Alarms] section of the [SpO
2
] menu, select the options described in the fol-
lowing table.
Options Description Settings*
Speed Selects the SpO
2
pleth waveform speed. 6.25 mm/s, 12.5 mm/s, 25
mm/s
Display SIQ
(Masimo only)
Selects whether or nor to show the Signal Index
Quality (SIQ) in the SpO
2
waveform area. The SIQ
wave indicates the confidence associated with
the saturation measurement and timing of the
pulse. Higher pulse is a better signal.
On, Off
* The factory default settings are in bold.