Operator's Manual
Monitoring ECG Relearning
TM80/TM70 Telemetry Monitor Operator’s Manual 7 - 35
7.10.2 Manually Initiating an ECG Relearning
1. Tap the QT digital area to enter the [QT/QTc] menu.
2. Tap [Relearn].
OR
1. On the main screen, tap the HR digital area or ECG waveform area to enter the
[ECG] menu.
2. In the [ECGParameter Setup] section, tap [
Arrhythmia].
3. Tap [Relearn].
CAUTION
• Initiate ECG relearning only during periods of normal rhythm and when
the ECG signal is relatively noise-free. If ECG learning takes place during
ventricular rhythm, the ectopics may be incorrectly learned as the nor-
mal QRS complex. This may result in missed detection of subsequent
events of V-Tach and V-Fib.