Operator's Manual
EMC EMC and Radio Regulatory Compliance
B - 2 TM80/TM70 Telemetry Monitor Operator’s Manual
B.1 EMC
The device meets the requirements of IEC 60601-1-2: 2014.
WARNING
• Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this device could result in increased
electromagnetic emissions or decreased electromagnetic immunity of
this device and result in improper operation.
• Use of this device adjacent to or stacked with other device should be
avoided because it could result in improper operation. If such use is nec-
essary, this device and the other device should be observed to verify
that they are operating normally.
• Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the this device, including cables specified
by the manufacturer. Otherwise, degradation of the performance of this
device could result.
• The non-ME EQUIPMENT (e.g. ITE) that is a part of an ME SYSTEM may be
disrupted by the electromagnetic interference of nearby equipment. It
may be necessary to take mitigation measures, such as re-orienting or
relocating the non-ME EQUIPMENT or shielding the location.
• This device is intended for use in professional healthcare facility envi-
ronment. If it is used in special environment, such as magnetic reso-
nance imaging environment, the equipment/system may be disrupted
by the operation of nearby equipment.
Guidance and Declaration - Electromagnetic Emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user of
the device should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment — guidance
RF emissions
CISPR 11
Group 1 The system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.