Operator's Manual
Measurement Specifications Product Specifications
A - 24 TM80/TM70 Telemetry Monitor Operator’s Manual
■ Modified Bland-Altman for SpO
2
- D-YS, OXI-P/I Sensors (No Motion): SaO
2
vs.
(SpO
2
- SaO
2
)
A.8.3 Resp
1. SpO
2
- SaO
2
(%) 2. SaO
2 (%)
3. Upper 95% LoA 4. Mean Bias 5. Lower 95% LoA
Technique Trans-thoracic impedance
Lead Lead II
Respiration Rate
Measurement range A
dult: 0 to 120 rpm
Pediatric: 0 to 150 rpm
Resolution 1 rpm
Accuracy 0 to 6 rpm: not specified
7 to 150 rpm: ±2 rpm or ±2% whichever is
greater