Operator's Manual
Product Specifications Measurement Specifications
TM80/TM70 Telemetry Monitor Operator’s Manual A - 23
A.8.2.6 Fitting Curve for Nellcor SpO
2
Sensors
■ The accuracy of Nellcor SpO
2
Sensors
■ Modified Bland-Altman for SpO
2
- DS-100A Sensor (No Motion): SaO
2
vs.
(SpO
2
- SaO
2
)
Sensor Motion # of Data Points
A
RMS
(%)
SpO
2
Acceptance
Criteria 70% - 100% (%)
Pass/
Fail
DS-100A
(PN:9000-10-05161 )
No 411 2.16 ≤3.0 Pass
D-YS (PN: 0010-10-
12476),
OXI-P/I (PN: 9000-10-
07308)
No 458 1.96 ≤3.5 Pass
1. SpO
2
- SaO
2
(%) 2. SaO
2 (%)
3. Upper 95% LoA 4. Mean Bias 5. Lower 95% LoA