Operator's Manual

Measurement Specifications Product Specifications
A - 20 TM80/TM70 Telemetry Monitor Operator’s Manual
A.8.2.4 Fitting Curve for Masimo SpO
2
Sensors
Table information for the plots below show ARMS values measured with Masimo SET
Oximetry Technology in a clinical study.
Adtx/Pdtx
SaO
2
range
Measured A
rms
Values
LNCS Adtx, LNCS Pdtx LNCS DCI, LNCS DCIP
70% to 100% ± 2% 2%
70% to 80% 1.55% 0.60%
80% to 90% 1.07% 0.54%
90% to 100% 1.64% 0.60%
PR
Measurement range
25 bpm to 240 bpm
Resolution
1 bpm
Accuracy
±3.0 bpm(measured without motion)
±5.0 bpm (measured with motion)
PI
Measurement range
0.02% to 20%
Least resolution
0.01, use three valid digits
Response Time
≤ 25 s (PR 75 bpm, average time 8 s, no disturbance, SpO
2
value rises from 50% to 100%)
≤ 20 s (SpO
2
value 98%, average time 8 s, no disturbance, PR value rises from 60 bpm to 150
bpm)