Operator's Manual
Measurement Specifications Product Specifications
A - 16 TM80/TM70 Telemetry Monitor Operator’s Manual
A.8.2 SpO
2
Time to alarm for tachycardia
Meets the requirements of IEC60601-2-27: Clause 201.7.9.2.9.101 b)
6).
Waveform
B1h - range: < 11 s
B1- range: < 11 s
B1d - range: < 11 s
B2h - range: < 11 s
B2- range: < 11 s
B2d- range: < 11 s
ST Segment Analysis
Measurement range
-2.0 mV to +2.0 mV RTI
Resolution
0.01 mV
Accuracy
-0.8 mV to +0.8 mV: ± 0.02 mV or ± 10%, whichever is greater
Beyond this range: Not specified
Arrhythmia Analysis
Arrhythmia analysis classifications
Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme
Brady, PVCs/min, Pauses/min, R on T, Run PVCs, Couplet, Multif. PVC,
PVC, Bigeminy, Trigeminy, Tachy, Brady, Pacer Not Pacing, Pacer Not
Capture, Missed Beat, Nonsus. Vtac, Vent. Rhythm, Pause, Irr.Rhythm,
Afib
QT Analysis
QTc formula
Bazett, Fridericia, Framingham, and Hodges
QT measurement range
[200, 800] ms
QT accuracy
[200, 800] ms: ± 30 ms
Beyond this range: Not specified
QT resolution
[200, 800] ms: 4 ms
Beyond this range: Not specified
QTc measurement range
[200, 800] ms
QTc resolution
[200, 800] ms: 1 ms
Beyond this range: Not specified
QT-HR measurement range
Adult: [15, 150] bpm
Pediatric: [15, 180] bpm