Operator's Manual

Wireless Specification Product Specifications
A - 10 TM80/TM70 Telemetry Monitor Operator’s Manual
A.6.2.2 Implemented Function
TM70 transmits waveforms, parameters, status and alarms of ECG, SPO2, RESP and
parameters, status and alarms of NIBP to the central station. The waveforms, parameters,
status and alarms displayed on central station are consistent with TM70.
TM70 can work well with APs which support POE, if the RSSI from APs is higher than -70
dBm.
A.6.2.3 Function Specifications
Modulation mode
GFSK
Operating frequency
608MHz(608~614MHz)
1.4GHz(1395~1400MHz and
1427~1432MHz)
Channel spacing
600kHz 800kHz or 600kHz
Wireless baud rate
(data rate)
Max 400kbps
Max 433kbps for 800kHz
channel spacing;
Max 400kbps for 600kHz
channel spacing.
Max. Output power
10 dBm
Receiver sensitivity
≤-90 dBm for AP70
≤-87 dBm for TM70
Data security
authentication: based on TLS ; encryption : AES-128bit
Radio Frequency Accuracy
during normal operation
(-50,+50) KHz relative to channel frequency
Occupied bandwidth as
defined by power in 99%
bandwidth
(-300,+300) KHz (-400,+400) KHz
WARNING
Do perform all network functions of data communication within an
enclosed network.
A.6.2 TM70 and AP70 WMTS Specifications
A.6.2.1 Technical Specification
Specifications 608M 1.4G