Operator's Manual
Classifications Product Specifications
A - 2 TM80/TM70 Telemetry Monitor Operator’s Manual
This chapter provides specifications of the TM80, TM70, AP70, SYNC70and central char-
ger, For specifications of BP10, refer to BP10 NIBP Module Operator’s Manual (P/N 046-
008269-00).
A.1 Classifications
The device is classified as follows according to IEC60601-1:
A.2 Environmental Specifications
Type of protection against electrical
shock
TM80/TM70: Class I, energized from an internal electrical power
source.
Central Charger: Class I.
AP70: Class II.
SYNC70: Class I.
Degree of protection against electrical
shock for the TM80/TM70
Type CF defibrillation proof for ECG and SpO
2
.
Mode of operation
Continuous
Degree of protection against harmful
ingress of water
TM80/TM70: IPX7
Central Charger: IPX0
AP70: IPX1
SYNC70:IPX0
AC70:IPX0
Degree of protection against hazards of
explosion
Not suitable: Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air with oxygen or nitrous
oxide.
Degree of protection against hazard of
dropping
No damage by dropping from a height of 1.5 m for the TM80/
TM70
WARNING
• The devices may not meet the performance specifications if stored or
used outside the specified temperature and humidity ranges.