Operator's Manual
Maintenance Verifying the ECG
TM80/TM70 Telemetry Monitor Operator’s Manual 17 - 5
17.6 Verifying the ECG
The ECG signal may be inaccurate due to hardware or software problems. As a result, the
ECG wave amplitude becomes greater or smaller.
To verify the ECG waveform amplitude:
1. In the main menu, tap [Maintenance].
2. Input the maintenance passcode.
3. Tap [Accept].
4. In the [Maintenance] menu, tap [General].
5. Enable [Calibrate ECG].
A square wave appears on the screen and the message [ECG Calibrating] is dis-
played.
6. Compare the amplitude of the square wave with the wave scale. The difference
should be within 5%.
7. After completing the verification, disable [Calibrate ECG].
17.6.1 NIBP Accuracy Test
The NIBP accuracy test is required at least twice a year or when you doubt the measured
NIBP result. Contact your service personnel to perform NIBP accuracy test.
17.6.2 NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required at
least twice a year or when you doubt the measured NIBP result. Contact your service per-
sonnel to perform NIBP leakage test.
17.6.3 Nurse Call Test
The nurse call test checks whether this functionality can work properly. Contact your ser-
vice personnel to perform the nurse call test.