Operator's Manual
Monitoring ECG Arrhythmia Monitoring
TM80/TM70 Telemetry Monitor Operator’s Manual 7 - 31
Irr Rhythm Configures arrhyth-
mia alarms.
On, Off
AF/Irr Rhy End Time: 0, 1 min, 2 min, 3 min, 4
min, 5 min, 10 min, 15 min, 30 min; th default
is 2 Minutes.
Alarm priority: Prompt, Low
A-Fib On, Off
AF/Irr Rhy End Time: 0, 1 min, 2 min, 3 min, 4
min, 5 min, 10 min, 15 min, 30 min; th default
is 2 Minutes.
Alarm priority: Prompt, Low, Med, High
PVCs/min On, Off
PVCs High: 1 to 100; the default is 10.
Alarm priority: Prompt, Low, Med, High
Pauses/min On, Off
Pauses High: 1 to 15; the default is 8.
Alarm priority: Prompt, Low, Med, High
Options Description Settings*
* The factory default settings are in bold.
WARNING
• If you switch off all arrhythmia alarms, the monitor will not alarm for any
arrhythmia event. This may result in a hazard to the patient. Always
keep the patient under close surveillance.