16 - 6 TM80/TM70 Telemetry Monitor Operator’s Manual
16.6 Disinfection
Disinfect the TM80/TM70 and accessories as required in your hospital’s servicing sched-
ule. Cleaning before disinfection is recommended.
Refer to the instructions for use delivered with the accessory to disinfect reusable ECG
leadwires and SpO
2
sensors.
The following table lists recommended disinfectants for disinfecting the TM80/TM70.
Product Name Product Type Active Ingredients
Isopropanol* Liquid Isopropanol 70%
1-Propanol* 1-Propanol 50%
CIDEX® OPA Solution Ortho-Phthalaldehyde 0.55%
Metrex CaviCide1
TM
Diisobutylphenoxyethoxyethyl dimethyl
benzyl ammonium chloride 0.28%,
Isopropanol 17.2%
Virex® II 256 Didecyl dimethyl ammonium chloride
8.704%,
n-Alkydimethyl benzyl ammonium
chloride 8.190%
Virex® TB n-Alkyl dimethyl benzyl ammonium
chlorides 0.105%,
n-Alkyl dimethyl ethylbenzyl ammonium
chlorides 0.105%
*Ethanol, 70% /
Hydrogen peroxide, 3% /
Sodium hypocholride (bleach)
solution, 10%
/
Isopropyl alcohol, 70% /
50% propyl alcohol (1-propyl
alcohol)
/
Super Sani-cloth 0.5% Quaternary ammonium chloride
and 55% Isopropyl alcohol
Rely+On
TM
Virkon® Powder *
(Used as 1% solution)
Powder Used as 1% solution
Biocidal active:
Pentapotassium bis
(peroxymonosulphate) bis
(sulphate)(500g/kg),
Contains dipotassium peroxodisulphate.