Operator's Manual

Cleaning and Disinfecting Cleaning the Battery and Battery Compartment
TM80/TM70 Telemetry Monitor Operator’s Manual 16 - 5
16.5 Cleaning the Battery and Battery Compartment
Clean the exterior surface of the lithium-ion battery and AA battery tray before
and after each new patient.
Wipe the lithium-ion battery, AA battery tray and the battery compartment using
a soft cloth with mild soap and water solution. Use caution to avoid the battery
contacts.
Wipe off all the cleaning solution with a dry cloth if necessary.
Allow the lithium-ion battery, AA battery tray and the battery compartment
completely dry before using.
CAUTION
Clean the cables carefully to avoid breaking internal wires by excessive
bending, strain, or flexing.
Do not autoclave, radiation or steam sterilize the cables and sensors.
Extended exposure to Ethylene Oxide gas may shorten life of the cables,
leading to poor signal quality.
NOTE
Refer to the individual instruction sheets that are packaged with the
accessories for additional information.
CAUTION
Never immerse the lithium-ion battery, AA battery tray and AA batteries
in any fluids. Do not clean them with harsh chemicals such as acetone or
diluted bleach.
As the battery compartment is not water proof, do not contact the
battery contacts when cleaning the battery compartment.
Do not hook the battery contacts in the battery compartment with other
objects. Otherwise, the battery contacts may be distorted.
Keep the battery compartment free away from any foreign objects.