Technical Alarm Messages at the TM80/TM70 Troubleshooting
15 - 10 TM80/TM70 Telemetry Monitor Operator’s Manual
Power Low Bat-
tery
Medium A The battery charge is low. Replace with new
batteries.
Critically
Low Bat-
tery
High A The battery charge is
almost depleted.
Battery
Mainte-
nance
Required
Medium A The lithium-ion battery is
aging.
Battery
Error
Medium A The lithium-ion battery
communication is error.
Battery
Type
Error
Medium A The battery contacts are
in bad contact.
System Device
Error
High A
The self-test of the main
board is error.
The self-test of the
parameter module or
bluetooth module are
error.
The parameter module
communication is error
or initialization is error.
Restart the TM80/
TM70. If the problem
persists, contact
your service person-
nel.
Real
Time
Clock
Error
High A The real time clock initial-
ization is error.
Restor-
ing Last
Defaults
Failed
Low C Restoring the last default
configuration is error.
Loading
Defaults
Failed
Low C Loading the default con-
figuration is error.
Measurement
Alarm
message
Alarm
level
Alarm
Indication
Possible cause Solution
* The asterisk (*) means the alarm level is configurable.
** XX is the ECG lead name.