Operator's Manual

Troubleshooting Technical Alarm Messages at the TM80/TM70
TM80/TM70 Telemetry Monitor Operator’s Manual 15 - 5
15.3 Technical Alarm Messages at the TM80/TM70
The following table lists the major technical alarm messages displayed in the message
area of the TM80/TM70.
The Alarm Indication column in the table below is capable of three different indication
types: A, B, C.
A: The alarm sound will be silenced, the alarm light will continue to indicate the
alarm, a will appear before the alarm message. The symbol appears on the
top of the main screen.
B: The technical alarm will be changed to the prompt message, it will not longer
make sound or be indicated by the alarm light.
C: The technical alarms are cleared, there will be no alarm indications.
NIBP NIBP-Sys
Extremely High
High The NIBP systolic
pressure has risen above
the SYS Desat alarm high
limit or fallen below the
SYS Desat alarm low
limit.
Check the patient’s con-
dition and check if the
alarm limit settings are
correct.
NIBP-Sys
Extremely Low
NIBP-Mean
Extremely High
The NIBP mean pressure
has risen above the MAP
Desat alarm high limit or
fallen below the MAP
Desat alarm low limit.
NIBP-Mean
Extremely Low
NIBP-Dia
Extremely High
The NIBP diastolic
pressure has risen above
the DIA Desat alarm
high limit or fallen below
the DIA Desat alarm low
limit.
NIBP-Dia
Extremely Low
Measurement Alarm message Alarm level Possible cause Solution
* The asterisk (*) means the alarm level is configurable.