Operator's Manual
Battery Maintaining the Rechargeable Lithium-ion Battery
TM80/TM70 Telemetry Monitor Operator’s Manual 14 - 7
14.8 Maintaining the Rechargeable Lithium-ion Battery
Take care of the rechargeable lithium-ion battery once you receive a new battery for use.
The following table describes the battery maintenance activities and recommended fre-
quency.
The lifetime of a lithium-ion battery depends on the frequency and duration of use. With
good maintenance, the useful life is approximately 500 complete charge-discharge
cycles. Experience indicates that the incidence of failure may increase with battery ser-
vice life due to the accumulated stresses of daily use. Therefore, Mindray strongly recom-
mend that lithium-ion battery should be replaced after two years or 500 complete
charge-discharge cycles. Using the outdated battery may cause the device abnormity
and unacceptable performance.
The age of a lithium-ion battery begins at the date of manufacture. The date of manufac-
ture is listed on the rear of the battery.
NOTE
• Replace the AA battery frame on the battery compartment after remov-
ing the AA batteries.
Activity Recommended Frequency
Visual inspection
Before installing a battery in the TM80/TM70.
Charge the battery Upon receipt, after use, a “Low Battery” or “Critically Low Bat-
tery” alarm occurs.
To optimize performance, a fully or almost fully discharged
battery must be charged immediately.
Clean the battery At each patient discharge, or in case that the battery is
exposed to contaminants.
Do not clean the battery connector during the cleaning.
Charge stored battery to at
least 40% of the battery
capacity.
Every six months if the TM80/TM70 is not in use for an
extended period of time.
Dispose of the battery When the “Battery Maintenance Required” alarm message
displays on the TM80/TM70.