Battery Charging the Rechargeable Lithium-ion Battery
TM80/TM70 Telemetry Monitor Operator’s Manual 14 - 5
2. Remove the lithium-ion battery pack or AA battery tray from the TM80/TM70.
14.6 Charging the Rechargeable Lithium-ion Battery
Use the central charger to charge the lithium-ion batteries. The central charger can
charge 10 lithium-ion batteries at one time. For details about the central charger, refer to
the BeneVison Central Charger Opeartor’s Manual (P/N 046-007059-00).
WARNING
Only use the specified central charger to charge to the lithium-ion bat-
teries designated by Mindray.
Only use the approved power cord with the grounded mains plug to
firmly connect the central charger to a grounded AC mains socket. Never
refit the mains plug to fit an ungrounded AC mains socket.
Do not use the Multiple Portable Socket Outlets (MPSO) or AC mains
extension cords. Use an IEC 60601-1 approved isolation / separation
transformer, otherwise, it may result in leakage current. Insure that the
sum of the individual ground leakage currents does not exceed the
allowable limits.
Do not place any shield object (such as cloth or paper) to cover the cen-
tral charger or batteries, and keep ventilated while charging the lithium-
ion batteries.
Do not connect other devices to the power supply system.
Do not use the central charger to charge the lithium-ion batteries in high
temperature above 40°C.