Operator's Manual
Introduction Battery
14 - 2 TM80/TM70 Telemetry Monitor Operator’s Manual
14.1 Introduction
The TM80/TM70 can be powered by a lithium-ion rechargeable battery or AA batteries.
This chapter provides instructions on how to use, maintain, and dispose of the batteries.
14.2 Safety
SE
WARNING
• Keep the batteries out of children’s reach.
• Keep the batteries in their original package until you are ready to use
them.
• Do install the lithium-ion battery pack or the AA battery frame to close
the battery compartment during defibrillation.
• Only use specified AA batteries or rechargeable lithium-ion battery to
power the TM80/TM70. Other power supply may cause damage to the
equipment or lead to body injury.
• While installing AA batteries, do not apply reverse polarity.
• Only use specified fresh AA batteries. Using other AA batteries can give
unacceptable performance.
• Do not mix old and new AA batteries.
• Only use specified rechargeable lithium-ion battery. Unspecified lith-
ium-ion battery can give unacceptable performance.
• Use caution when handling the rechargeable lithium-ion battery. Mis-
use or abuse may cause bodily injury or device damage.
◆ Do not short circuit. Take care that the terminals do not contact
metal or other conductive materials during transport and storage.
◆ Do not crush, drop or puncture the battery. Mechanical abuse can
lead to internal damage and internal short circuits. If a battery has
been dropped or banged against a hard surface, whether damage
is externally visible or not, remove the battery from use and dis-
pose of it properly.
◆ Do not incinerate batteries or expose them to temperatures above
60°C (140°F).