7 - 28 TM80/TM70 Telemetry Monitor Operator’s Manual
Options Description Settings*
Asystole Configures arrhyth-
mia alarms.
On, Off
Asystole Delay: 3 seconds to 10 seconds
The default is 5 Seconds.
Alarm priority: High
V-Fib/V-Tach On, Off
Alarm priority: High
V-Tach On, Off
V-Tach PVCs: 3 beats to 99 beats; the default
is 6 beats.
V-Tach Rate:100 bpm~200 bpm; the default
is 130 bpm.
Alarm priority: High
* The factory default settings are in bold.