Operator's Manual

Setting CMS Disconnect Alarm Configuring the TM80/TM70
13 - 16 TM80/TM70 Telemetry Monitor Operator’s Manual
13.3 Setting CMS Disconnect Alarm
In the [Maintenance] menu, you can set whether to enable [CMS Disconnect Alarm]
This option is enabled by default.
When it is enabled, after the TM80/TM70 is disconnected from the CMS:
The TM80/TM70’s display is turned on automatically.
The TM80/TM70 issues a low-pitched beep, its alarm light indicator flashes in
cyan, and the “No CMS” message is displayed in the prompt message area of
the TM80/TM70.
When it is disabled, after the TM80/TM70 is disconnected from the CMS:
The TM80/TM70’s display is not turned on automatically. But you can turn on
the display manually.
The TM80/TM70 does not issue a low-pitched beep and its alarm light indi-
cator does not flash. But the “No CMS” message is still displayed in the
prompt message area of the TM80/TM70.
13.3.1 Demo Mode
Allows an authorized user to choose a demonstration mode for in-servicing staff or test-
ing product features.
13.3.2 Service Menu
Allows an authorized user access to the passcode protected Service menu.
13.4 Configuring WMTS Setting
You can set up AP Code and WMTS AP Band for TM70 in the [WMTS Setting] menu.Fol-
low this procedure:
1. In the [Maintenance] menu, tap [Service].
2. Input the password and tap [Accept].
3. Tap [WMTS Setting] and select the options described in the following table:.
WARNING
When [CMS Disconnect Alarm] is disabled, the medical staff should pay
more attention to the patient’s status and check the TM80/TM70’s net-
work connection status at the CMS. When the offline alarm is displayed
on the CMS, medical staff should check the patient ‘s status immedi-
ately.