Operator's Manual

Maintenance Menu Configuring the TM80/TM70
13 - 14 TM80/TM70 Telemetry Monitor Operator’s Manual
menu. On the TM80/TM70, the status for the external device will display the
“Complete” message.
On the external device if the import fails, the external device will display the
“Import attempt failed.” message. On the TM80/TM70, the status for the failed
external device will display the “Failure” message and remain in the list.
The user has two options when a failure occurs:
To stop the import from the external device, tap the icon to exit the
[Import Device Settings].
To retry the import from the TM80/TM70 tap the external device which is
displaying the [Failure] message, tap the [Retry] button to restart the trans-
fer.
You may need to repeat the retry operation several times until the transfer is
successful.
13.2.8 Screen Lock Menu
13.2.8.1 Understanding the Screen Lock Mode
There are two modes of being able to lock the screen to assist in preventing unautho-
rized use. Each mode allows the user access to certain features of the product without
entering a passcode. When the correct passcode is entered, all features are available.
Locked Mode Features:
Main screen and main menu are not accessible without passcode entry.
The message area is still viewable.
Hardkeys are enabled.
View Only Features:
Upon powering up, the main screen will be displayed after the new patient
choice is made.
Main Screen is accessible without passcode entry.
The [System Info] menu is accessible by tapping the battery symbol on the
main screen.
The [Lead Placement] menu is accessible by tapping anECG Sensor Off”
message in the alarm area.
Ability to change display orientation.
Hardkeys are enabled.