Operator's Manual

Configuring the TM80/TM70 Maintenance Menu
TM80/TM70 Telemetry Monitor Operator’s Manual 13 - 13
3. Tap to exit the [Defaults] menu.
13.2.7 Transferring a Configuration
The TM80/TM70 is capable of configuring multiple devices through one export opera-
tion via a wireless interface.
To transfer the configuration from the TM80/TM70 to an external device:
1. In the [Maintenance] menu of an external device, tap [Defaults] [Import
Device Settings] to enter the settings import mode.
2. In the [Maintenance] menu of the TM80/TM70 unit, tap [Defaults] [Export
Device Settings] to enter the [Export Device Settings] menu.
The discovered external devices are listed in the [Export Device Settings] menu.
3. Select the desired external devices by tapping the check box.
4. Tap [Export] to start exporting the TM80s configuration.
The selected external devices screen will shortly display the “Downloading device
settings” message.
On the external devices if an import is successful, the external device will display
the “Download complete.” message, and after 10 seconds return to the [Defaults]
Export Device Settings Copies the current settings to the external device.
Import Device Settings Copies the settings from the external device to the TM80/TM70 unit. Refer to
"Transferring a Configuration" on page 13 - 13 for details.
Restore Factory Defaults Allows an authorized user to reestablish the original database power up
settings to factory default values.
Options Description
* The factory default settings are in bold.
WARNING
Do not power off the devices during the download process.
Configurations can only be transferred between telemetry devices of
the same model. You cannot transfer configuration from TM80 to TM70,
or vice versa.