Operator's Manual
Monitoring ECG Arrhythmia Monitoring
TM80/TM70 Telemetry Monitor Operator’s Manual 7 - 27
Note: N: normal beat; V: ventricular beat
7.9.3 Changing Arrhythmia Settings
7.9.3.1 Changing Arrhythmia Alarm Settings
1. On the main screen, tap the HR digital area or ECG waveform area to enter the
[ECG] menu.
2. In the [ECGParameter Setup] section, tap [
Arrhythmia].
3. Tap [Arrhythmia Alarm Setup]. The [Arrhythmia Alarm Setup] menu is dis-
played.
4. In the [Setup] section, select the desired item:
◆ [All On]:switches on all arrhythmia alarms
◆ [All Off]:switches off all arrhythmia alarms
◆ [Lethals Only]: switches on all the lethal arrhythmia alarms only. Other
types of arrhythmia alarms are off.
5. In the [Alarms] section, configure the options described in the following table.
Afib P wave is absent and normal beat RR intervals are irregular.
PVCs/min PVCs/min exceeds high limit.
Pauses/min Pauses/min exceeds high limit.
Irr. Rhythm End Irregular rhythm no longer detected for the irregular rhythm end
delay time.
Afib End Atrial fibrillation no longer detected for the Afib end delay time.