Operator's Manual

Table Of Contents
7 - 14 BeneVision N1 Patient Monitor Operator’s Manual
7.12 Managing Alarms from Remote Devices (only available for the
independent external display)
You can view and manage the alarms coming from remote devices, for example bedside monitors and telemetry
monitors on your monitor.
7.12.1 Setting the Tone Pattern for Alarms from Remote Devices
The monitor provides the same alarm tone pattern for the remote device alarms as those for your monitor
alarms. For more information on alarm tone pattern, see 7.3.3 Alarm Indicators.
To set the alarm tone pattern for the remote device alarms, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Alarm tab select the Remote View tab.
3. Set Alarm Sound to ISO, Mode 1 or Mode 2.
If your monitor and remote devices have alarms at the same time, the monitor presents alarms according to the
following rules:
If the alarm tone pattern for your monitor is same as that for the remote devices, the monitor prioritizes all
the alarms and sounds the highest alarm tone.
If the alarm tone pattern for your monitor is differently from that for the remote devices, the monitor
sounds the alarms in their corresponding tone.
The monitor provides the same alarm tone pattern for the remote device alarms as those for your monitor
alarms by default. You can change the alarm tone pattern of the remote monitor; see 37.3.4.5Setting the Tone
Pattern for Alarms from Remote Devices
7.12.2 Selecting the Alarm Reminder for Remote Devices
You can configure what alarm indicators are necessary for the remote devices. To do so, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Alarm tab Remote View tab.
3. Select Alarm Reminder:
Visible + Audible: the monitor provides visual alarm indication, and continuous audible alarm
indication if the alarm persists at the remote device.
Visible + Single Tone: the monitor provides visual alarm indication, and a single tone when the alarm
occurs at the remote device.
Visible Only: the monitor only provides visual alarm indication.
7.12.3 Presenting Alarm Sound for Remote Devices as per Alarm Priority
You can configure what priority of remote device alarms are presented for audible notification. To do so, follow
this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Alarm tab Remote View tab.
3. Select Alarm Priority:
All: the monitor sounds if an alarm occurs.
High&Med: the monitor sounds if a high or medium priority alarm occurs.
High Only: the monitor sounds only if a high priority alarm occurs.