7 - 8 BeneVision N1 Patient Monitor Operator’s Manual
7.6.5 Adjusting the Alarm Light Brightness
To adjust the alarm light brightness, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Alarm tab Other tab.
3. From the Alarm Light Brightness block, set Primary Display. If you set alarm light brightness to Auto, the
monitor automatically adjusts the alarm light brightness according to the ambient light. The stronger the
ambient light is, the brighter the alarm light is.
For N22/N19, if you are using a secondary display, set the Secondary Display too.Restoring the Default Alarm
Settings
To reset all alarm settings to the defaults, follow this procedure:
1. Select the Main Menu quick key from the Alarm column select Setup.
2. Select the Limits tab.
3. On the Limits page, select Defaults at the left bottom.
7.6.6 Setting the Length of Printing Duration On Alarm
You can define the length of printed waveforms when an alarm is triggered. To do so, follow this procedure:
1. Access the Setup page in either of the following ways:
Select the Alarm Setup quick key select the Setup tab.
Select the Main Menu quick key from the Alarm column select Setup.
2. Select the Setup tab.
3. Set Printing Duration On Alarm.
7.6.7 Setting the Switch of the CMS and eGateway Disconnection Alarm
You can choose whether to issue an alarm when the monitor is not connected or disconnected from the CMS/
eGateway. To do so, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Alarm tab Other tab.
3. Switch on or off CMS/eGW Disconnected Alarm. If CMS/eGW Disconnected Alarm is switched off, the
“Offline” alarm is not presented when the monitor is not connected or disconnected from the CMS/eGate-
way.
7.6.8 Setting the Priority of the CMS and eGateway Disconnection Alarm
To set the priority of the CMS and eGateway disconnection alarm, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Alarm tab Other tab.
3. Set CMS/eGW Disconnected.
7.6.9 Setting the Switch of the SpO
2
Desat Alarm Off
You can choose whether switching off the SpO
2
Desat alarm is permissible or not. To do so, follow this
procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .