BeneVision N1 Patient Monitor Operator’s Manual 7 - 7
7.6.4 Setting the Alarm Delay Time
For continuously measured parameters, you can set the alarm delay time. If the alarm condition is resolved
within the delay time, the monitor does not present the alarm.
To set the alarm delay time, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Alarm tab Other tab.
3. Set Alarm Delay.
The setting of Alarm Delay is not applied to the zero RR alarms and the ST alarms. You can set Zero RR Delay
and ST Alarm Delay separately.
7.6.4.1 Setting the Zero RR Delay Time
To set the Zero RR delay time, follow this procedure:
1. Select the Main Menu quick key from the Alarm column select Setup.
2. Select Zero RR Delay to set the zero RR delay time.
7.6.4.2 Setting the ST Alarm Delay Time
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Alarm tab Other tab.
3. Set ST Alarm Delay.
IBP: CVP/
LAP/
RAP/
UVP/P1-
P4
(Venous
pressure)
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 40
CO
2
EtCO
2
(mmHg)
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
Same as the
measurement range
33 to 35: 29 33 to 35: 29 33 to 35: 41 33 to 35: 41 Same as the
measurement range
36 to 45:
(etCO
2
- 6)
36 to 45:
(etCO
2
- 6)
36 to 45:
(etCO
2
+ 6)
36 to 45:
(etCO
2
+ 6)
Same as the
measurement range
46 to 48: 39 46 to 48: 39 46 to 48: 51 46 to 48: 51 Same as the
measurement range
>48: remains
the same
>48: remains
the same
>48: remains
the same
>48: remains
the same
Same as the
measurement range
FiCO
2
None None Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
awRR
(rpm)
awRR × 0.5 or
6 (whichever
is greater)
(awRR - 10) or
30 (whichever
is greater)
awRR × 1.5 or
30 (whichever
is smaller)
(awRR+25) or
85 rpm
(whichever is
smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
Module Parameter
Lower Limit Upper Limit
Auto Limit Range
Adult/
pediatric
Neonate
Adult/
pediatric
Neonate