6 - 4 BeneVision N1 Patient Monitor Operator’s Manual
6.7.1.3 Adding a Bed
You need to add the desired remote devices, and then the alarms from these devices can be watched on your
monitor. To add a remote device, follow this procedure:
1. Enter the Select Bed window. To do so, choose either of the following ways:
In the Remote View window, select Select Bed. For more information, see 6.7.1.1 Entering the Remote
View Window for entering the Remote View window.
Select the icon at the alarm watch tile if the tile is configured to display on the main screen.
2. In the Select Bed window, select a desired department. All the beds under this department will be listed.
3. Select a desired tile at the A-W1, or A-W2 areas and then select a bed from the bed list. The selected bed
will appear in the tile.
NOTE
The added bed is indicated by a check mark at the right of the bed list.
6.7.1.4 Removing a Bed
If you do not want to monitor a remote device any longer, you can remove it. To remove a remote device, follow
this procedure:
1. Enter the Select Bed window. Choose either of the following ways:
In the Remote View window, select Select Bed. For more information, see 6.7.1.1 Entering the Remote
View Window for entering the Remote View window.
Select the icon in the alarm watch tile if the tile is configured to display on the main screen.
2. In the Select Bed window, select a bed at the A-W1 or A-W2 area, and then select Clear Bed. If you want
remove all beds, select Clear All Beds.
6.7.1.5 Displaying the Main Bed
In the Remote View window, you can select a bed at the alarm watch area, then the main body of the Remote
View window will display the real time monitoring screen of the device.
6.7.1.6 Saving a Manual Event
You can initiate a manual event by selecting Manual Event in the Remote View window.
The manual event stores in the event review of the corresponding remote device.
6.7.1.7 Resetting Alarms for Remote Devices
You can reset the alarms on the remote devices by selecting Alarm Reset in the Remote View window. This
function needs to be enabled. To enable this function, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Alarm tab Remote View tab.
3. Switch on Reset Remote Bed’s Alarms. Then the Alarm Reset button appears on the bottom left of the
Remote View window.
4. To reset remote device alarms, select Alarm Reset.
NOTE
You can reset remote device alarms only if the Alarm Reset by Other Bed switch is on at the remote
devices.
(2) Main body
Display the patient’s information, alarm status and messages, waveforms, measurements, etc. of the selected bed.
This bed is called main bed.