Operator's Manual

Table Of Contents
4 - 4 BeneVision N1 Patient Monitor Operator’s Manual
4.3.8 Setting Monitor Location
If your monitor is installed at a fixed position, or you do not need to frequently change the room number and
bed number, you can set Monitor Location to Fixed. To set the monitor location, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Device Location tab.
3. Set Location.
Fixed: the Patient Management menu displays Bed No. and Room No., but you cannot change
them.
Unfixed: you can change Bed No. and Room No. from the Patient Management menu.
4. Input room number and bed number.
NOTE
If Location is set to Unfixed, Bed No. and Room No. are cleared each time you discharge a patient.
4.3.9 Setting Patient Location
To set patient location, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Patient Management tab.
3. Select the Location tab.
4. Input patient location as desired.
4.3.10 Hiding Patient Name
To hide patient name, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Patient Management tab.
3. Select the Display tab.
4. Set the following switches as desired:
Switch off Primary Screen Display Full Name: does not display patient name in the patient
information area on the primary display.
Switch off Remote View Display Full Name: does not display patient name in the patient
information area on the remote monitors when this monitor is viewed by other monitors.
Switch off Remote View Bedlist Display Full Name: does not display patient name in bed list on the
remote monitors when this monitor is viewed by other monitors.
4.4 Transferring Patient Data
You can transfer the patient data between N1 and the host monitor without re-entering the patient
demographic information or changing the settings. Transferring of patient data enables you to understand the
patient’s history condition. The patient data that can be transferred includes: patient demographics, trend data,
event data, full disclosure waveform and parameters settings. The trends data and event data from the
parameter modules of the host monitor can also be transferred.
For detailed information about patient data transfer, refer to the user manual of the host monitor. For the
connection of N1 and the host monitor, refer to section 2.8 N1 in Use with the Host Monitor.
WARNING
Do not discharge a patient before the patient is successfully transferred.