Operator's Manual

Table Of Contents
BeneVision N1 Patient Monitor Operator’s Manual 3 - 13
3. Set Enter Outdoor Mode to Manual or Auto.
Manual: The monitor enters the outdoor mode by manually selecting the Main Menu quick key
from the Display column select Enter Outdoor Mode.
Auto: The monitor enters the outdoor mode automatically if the strength of ambient light is greater
than the threshold.
3.7.6.2 Exiting the Outdoor Mode
When Enter Outdoor Mode is set to Manual, select the Main Menu quick key from the Display column
select Exit Outdoor Mode.
The monitor automatically exits the outdoor mode in the following situation:
The monitor is connected to a host monitor.
The strength of ambient light is lower than the threshold when Enter Outdoor Mode is set to Auto.
3.8 Configuring Your Monitor
Configure your monitor before putting it in use.
3.8.1 Selecting the Language
To set the user interface (UI) language, follow this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Other tab.
3. Set Language.
3.8.2 Setting the Date and Time
To set the system time, follow this procedure:
1. Select the Main Menu quick key from the System column select Time.
2. Set Date and Time.
3. Set Date Format.
4. If you want to use the 12-hour mode, switch off 24 Hour Time.
5. If you want to use daylight saving time, switch on Daylight Saving Time. You can manually switch on or off
the daylight saving time only when the auto daylight saving time function is disabled. For more
information, see 3.8.3Enabling Auto Daylight Saving Time for details.
If your monitor is connected to a central monitoring system (CMS) or hospital clinical system (HIS), the date and
time are automatically taken from the CMS. In this case, you cannot change the date and time from your
monitor.
CAUTION
Changing the date and time affects the storage of trends and events and may result in loss of data.
3.8.3 Enabling Auto Daylight Saving Time
By default, you need to manually enable the daylight saving time. To auto start the daylight saving time, follow
this procedure:
1. Select the Main Menu quick key from the System column select Maintenance input the required
password select .
2. Select the Time tab.
3. Switch on Auto Daylight Saving Time.
4. Adjust daylight saving time settings as necessary.